Details of EMC TestLab facilities

Overview of measuring capabilities of the EMC TestLab

Facilities

Equipment

Frequency / Levels

Screened rooms

Semi-Anechoic room: SAR10: 20 x 12 x 9 m,
Turntable = 4 m, 4000 kg

9 kHz – 26.5 GHz

Semi-Anechoic room: SAR3: 9 x 6 x 6 m,
Turntable = 2 m, 1000 kg

9 kHz- 18 GHz

Semi-Anechoic room: SAR3/FAR3: 9 x 6 x 6 m,
Turntable = 2 m, 1000 kg

9 kHz - 18 GHz

Full-Anechoic room: FAR3: 7 x 3 x 3 m approx.,
Turntable = 2 m, 1000 kg

80 MHz - 18 GHz

Reverberation Chamber MSC14: 14 x 7 x 6 m
Sizes (LxWxH) lined with ferrite tiles and RF absorbers

80 MHz - 40 GHz

Compact FAR: 0.9 x 0.9 x 2 m

800 MHZ - 40 GHz

14 shielded rooms sizes from approximately 2 x 3 x 2.5 m, 3.5 x 5 x 2.5 m, to 5 x 6.5 x 2.5 m and 3.5 x 8.5 x 2.5 m

9 kHz - 18 GHz

Emission

Radiated emission with antennas in Semi-Anechoic room or Reverberation Chamber

9 kHz - 40GHz

Radiated emission in GTEM cell

9 kHz - 5GHz

Automotive Strip line

10 kHz - 1000 MHz

Conducted emission with Artificial Mains Network or Telecom Terminal Network

9 kHz - 30 MHz

Conducted emission with sliding run absorbing clamp and combining network (antenna terminal disturbance)

30 MHz - 1 GHz

Magnetic fields with loop sensor

50 Hz - 50 kHz

Harmonics/voltage fluctuations test system

3 x 2000 VA

Immunity

Radiated immunity with antennas in anechoic chamber

150 kHz - 26 MHz: 10 V/m @ 1 m.

26 MHz - 4000 MHz:
10 V/m @ 3 m.

Radiated immunity with antennas in a mode stirred chamber

80 MHz - 40 GHz

Radiated immunity in open strip-line TEM device:
(Transversal Electro-Magnetic cell), height = 0.8 m

150 kHz - 150 MHz: 10 V/m.

Radiated immunity in GTEM cell

DC-18 HGz, 100W

Radiated immunity in Compact FAR

800MHz - 40 GHz;
> 30 V/m @ 0.5 m.

Automotive Strip line

10 kHz - 1000 MHz:
200 V/m.

Conducted immunity with BCI test method

10 KHz - 1000 MHz:
200 V/m.

Conducted immunity with CDNs:
(Coupling/Decoupling networks)

150 kHz - 230 MHz:
10 Vemf on CDN input

ESD (Electro-Static Discharge) test set-up:

30 kV air and 15 kV contact discharge

Transient burst simulator:

3 phase, 4.4 kV,
5/50 ns

Surge simulator:

3 phase, 4 kV,
 1.2/50 ms

Magnetic field immunity test:

- with (Helmholtz) coil

- with near field coils

50 Hz - 10 kHz:

10 A/m

1000 A/m

Power fail simulator: voltage dips, interruptions and variations.

1 phase, 10 A

Automotive voltage according to ISO 7637

0 – 60 V battery voltage